AirQlean H™ - a ceiling mounted air filtration system
Open the catalog to page 1User's manual - AirQlean H” © Copyright 2014 QleanAir Scandinavia — 2
Open the catalog to page 2User's manual - AirQlean H” SECTION 1 Safety information 1.1. Introduction This chapter contains safety information. Main voltage powers AirQlean H™. Before you operate or do maintenance on the unit, you must know the hazards described in this manual. Obey the instructions in this manual to avoid personal injuries or damage to the equipment. This user documentation contains WARNINGS, CAUTIONS and NOTICES that concerns a safe use of the product. WARNING, highlight instructions that must be obeyed to avoid personal injury. Do not proceed until all stated conditions are clearly understood and...
Open the catalog to page 3CAUTIONS highlight instructions that must be obeyed to avoid damage to the product or other equipment. Do not proceed until all stated conditions are met and clearly understood. Always make sure that the supply voltage at the wall outlet is the same as the marking on AirQlean H, before connecting the power cord. o Make sure that AirQlean H is not exposed to water. o AirQlean H must be operated in a normal room environment. o Make sure that no sharp objects come in contact with the filters. The filters may be damaged. AirQlean H will not operate correctly if the filters are damaged. o Notice...
Open the catalog to page 4User's manual - AirQlean HT’ 2.1. Introduction AirQlean H is a ceiling mounted air filtration system. It cleans the air from gases and particles. AirQlean H uses a system of high performance HEPA and carbon filters. AirQlean H cleans up to 800 m3 airs per hour. The filtered air is 100 % clean. Only operate AirQlean H indoors in a temperature-controlled environment with normal humidity. See the chapter Installation for more details. The illustration below shows AirQlean H. © Copyright 2014 QleanAir Scandinavia - 5
Open the catalog to page 5User's manual - AirQlean H” 2.3. Dimensions The illustration and the table below show the dimensions of AirQlean H. © Copyright 2014 QleanAir Scandinavia - 6
Open the catalog to page 6User's manual - AirQlean H” 2.4. Illustrations of the main parts The illustrations below show the locations of the main parts of AirQlean H. © Copyright 2014 QleanAir Scandinavia — 7
Open the catalog to page 7User's manual - AirQlean H” © Copyright 2014 QleanAir Scandinavia — 8
Open the catalog to page 8User's manual - AirQlean H" 2.5. Control panel overview The Control panel is located under the Maintenance hatch. The Maintenance indicators and the Power switch can be accessed both with the Maintenance hatch open and closed. You must open the hatch to access the other controls. Control panel overview © Copyright 2014 QleanAir Scandinavia - 9
Open the catalog to page 9User's manual - AirQlean H” 2.6. Label The illustration below shows the AirQlean H label. It is positioned visible next to the power switch. © Copyright 2014 QleanAir Scandinavia — 10
Open the catalog to page 10User's manual - AirQlean H” SECTION 3Installation 3.1. Introduction In this document you will get all necessary information regarding the preparations you as a customer need to make before we can install our unit. If you have questions after reading the documents, contact QleanAir Scandinavia. 3.2. Site requirement The AirQlean H must be operated in a temperature-controlled environment with normal humidity. Make sure that: • Relative humidity does not exceed 60 %. • Temperature does not exceed 40°C. 3.3. Power requirements The table below shows the power requirements. Read and follow the...
Open the catalog to page 11Placement It is recommended to place the AirQlean H in the centre of the designated area. Installation requirement Minimum distance to wall, obstacle etc. Front side (with lock see 2.4.1) Left, right and backside (see 2.4.1) With Silencer – only right side The air outlet of the AirQlean H is located at the same side as the power socket (right side) and is recommended to be placed away from the entrance of the designated area. AirQlean H will not have any impact on the ventilation of the room. QleanAir Scandinavia recommends a negative pressure in the smoking room of 5 Pa and minimum 1 Pa. ©...
Open the catalog to page 12User's manual - AirQlean H” 3.6. Mounting plate You will get a mounting plate sent to you before the installation. The mounting plate shall be fixed to the ceiling before the technicians from QleanAir arrives. You, as customer are responsible to install the mounting plate in a secure way. • The customer is responsible for fixation of the Mounting plate. • The Mounting plate has a left and a right side, make sure it is correct fixated (see requirements 3.5). • The Mounting plate must be fixated in a secure way and according to local legislation. • It is recommended use at least 8 holes to...
Open the catalog to page 13User's manual - AirQlean H” 3.8. Flush installations We recommend that the unit is hanging under the ceiling. By recirculation the cleaned air you get a faster cleaning of the room. • If you want the unit’s lower line flush to (be in line with) the ceiling you must prepare the opening in the ceiling. • Notice that it is not possible to install the unit flush with silencer included. In that case, the unit must be hanging under the ceiling. • Notice, when installing the unit flush: inspect the space above the ceiling for channels and piping etc. to other nearby rooms. • Notice, the sound from...
Open the catalog to page 14User's manual - AirQlean H” © Copyright 2014 QleanAir Scandinavia - 15
Open the catalog to page 15Operation 1. Push the Power switch ( 8 ) to turn AirQlean H on or off. 2. When motion sensor ( 9 ) detects movement it starts the selected operation mode. 3. 20 minutes after last detected movement the unit goes to selected standby mode. 4. Clean the system regularly with a soft damp tissue and, if necessary, a mild detergent. This system is NOT intended for use by persons (including children) with reduced physical, sensory or mental capabilities, or lack of experience and knowledge, unless they have been given supervision or instructions concerning use of the appliance by a person...
Open the catalog to page 16