PREPARATION ON SITE 39 >
Wall-mounting plates
Installation instructions
Installation formulasGL WA >
– Mark centreline
– Mark and make wall opening relative to edge of medical supply unit - minimum front wall clearance is 10 mm– Place rear-end feed socket in position – order no. 20 961 303
– Insert electrical cables for sockets, nurse call, telephone
etc. as per production draw- ing allow excess length of at least 600 mm– Install medical gases as per drawing, observe correct sequence!
815 to 1000 mm GL-8041001 to 1609 mm GL-965
1610 to 2408 mm GL-1608
2409 to 4001 mm GL-(2401)/2
4002 to 5587 mm GL-(3194)/3
5588 to 6377 mm GL-(3987)/4 Non-sectionalised wall plates available on request Abbreviations GL = Total length
MA = Mounting centreline
OM = Upper mounting centreline
UM = Lower mounting centreline
WPA = Wall-plate clearance Art. Nr. 20 961 312Art. Nr. 20 961 334Art. Nr. 20 961 328 79311 185 WPA >
OMUM MA < 1000 GL 79311 800 WPA 15011 >
MA
OMUM 7474
7474 1001 to 1610 GL 7931111 800 WPA 793 >
MA 7474 1611 to 2410 GL >
UMOM 1315 OKFFB