PREPARATION ON SITE 33
Recessed into wall Surface-mounted on wall
– Mark out edges of medical supply unit– Mark out wall opening relative to edges of medical supply unit and make wall opening– Fit the mounting frame– Lead in electric cables, allow sufficient extra length to reach central connection area of
medical supply unit– Note: Where CONBOARD is recessed into the wall, patient call systems are connected
directly to component with-
out cutting; minimum excess
length of leads 500 mm – Mark out feed-in areas of medical supply unit starting
from bottom edge and
centreline– Ensure electric cables are long enough to reach central connection area of medical
supply unit– Note: Where CONBOARD is surface-mounted on the wall, patient call systems are connected directly to com-
ponent without cutting; mini-
mum excess length of leads
500 mm– Transfer fixing points
– Drill fixing holes and fit wall plugs General information concerning installation Installation must take place in
accordance with instructions
and must only be carried out by
expert skilled personnel.
600
GO GO >
80
80 80 385 360 80 >
625 300GAS
GU GU >
MA
1360 >
= possible feed-in area
1000 >
AbbreviationsGO = Top side of unit
GU = Underside of unit
MA = Mounting centreline
OKFFB = Top edge of finished floor
Recommended installation height OKFFB OKFFB >